Skip to main content
. 2020 Dec 7:lmaa104. doi: 10.1093/labmed/lmaa104

Table 2.

Performance Characteristics of RAD Test for the Presence of SARS-CoV-2 in 188 RT-PCR-Positive Specimens

All Specimens High-Viral-Load Specimens
Ct Value Number of Specimens Ct Value Number of Specimens
Mean Range Tested Positive Sensitivity Mean Range Tested Positive Sensitivity
22 15.8–32.3 188 81 43.1% 18 15.8–25.5 76 71 93.4%

Ct, cycle threshold; RAD, rapid antigen detection; RT-PCR, reverse-transcription polymerase chain reaction. High-viral-load specimens are those with Ct values <25.5 for SARS-CoV-2–specific RT-PCR, and normal viral load specimens are those with Ct values >28.0.