Skip to main content
. 2021 Jan 6;2021:6661744. doi: 10.1155/2021/6661744

Table 3.

Robustness of the method for potentiometric and complexometric titration.

Potentiometric titration Complexometric titration
Critical value 0.0022 0.0038
HCl volume 0.0006 NR
Heating temperature  NR 0.0081
pH NR 0.0027
Rest time 0.0016 0.0126
Sample weight NR 0.0223
Starring time 0.0017 0.0018
Starring velocity 0.0014 NR
Titrant addition rate 0.0017 NR

NR: no response.