Comparison between the mean values measured by the ANTERION and IOLMaster 500.

Parameter ANTERION IOLMaster 500 95% LoA p value ICC
Mean ± SD Mean ± SD
AL (mm) 23.62 ± 1.63 23.63 ± 1.62* – 0.06 to + 0.05  < 0.0001 1.000
ACD (mm) 3.21 ± 0.43 3.17 ± 0.43 – 0.50 to + 0.57 n.s 0.888
Sim-K (D) 43.85 ± 1.53 43.84 ± 1.58 – 0.68 to + 0.70 n.s 0.987
CD (mm) 11.83 ± 0.46 12.04 ± 0.39 – 0.72 to + 0.31  < 0.0001 0.839

SD standard deviation, LoA limits of agreement, ICC Intraclass correlation coefficient (between ANTERION and IOLMaster 500), AL axial length, ACD anterior chamber depth, Sim-K simulated keratometry, CD corneal diameter.