Table 4.
Parameter | ANTERION | IOLMaster 500 | 95% LoA | p value | ICC |
---|---|---|---|---|---|
Mean ± SD | Mean ± SD | ||||
AL (mm) | 23.62 ± 1.63 | 23.63 ± 1.62* | – 0.06 to + 0.05 | < 0.0001 | 1.000 |
ACD (mm) | 3.21 ± 0.43 | 3.17 ± 0.43 | – 0.50 to + 0.57 | n.s | 0.888 |
Sim-K (D) | 43.85 ± 1.53 | 43.84 ± 1.58 | – 0.68 to + 0.70 | n.s | 0.987 |
CD (mm) | 11.83 ± 0.46 | 12.04 ± 0.39 | – 0.72 to + 0.31 | < 0.0001 | 0.839 |
SD standard deviation, LoA limits of agreement, ICC Intraclass correlation coefficient (between ANTERION and IOLMaster 500), AL axial length, ACD anterior chamber depth, Sim-K simulated keratometry, CD corneal diameter.