Table 2.

Experimentally measured SAR (mean ± SD) using four optical E-field sensors at Siemens 3 T and two of same model Philips 3 T systems the average of the local SAR measurements with and without artificial hip joints at all locations from the head to the tail of the implants.

Image sequence Siemens 3 T Philips 3Ta Philips 3 Tb
T1w TSE T1 TIRM T2w TSE T1w TSE T1w SPIR T2w TSE T1w TSE T1w SPIR T2w TSE
Averaged local SAR with (without) implant (W/kg) 12.09 (4.19) 13.92 (4.19) 11.73 (4.19) 13.61 (4.07) 12.41 (4.07) 11.93 (4.07) 13.45 (4.24) 11.94 (4.24) 11.89 (4.24)
SAR at OEFS #1 18.32 ± 1.47 19.14 ± 1.35 16.77 ± 1.34 19.02 ± 1.45 18.51 ± 1.38 23.35 ± 1.78 18.16 ± 1.45 19.37 ± 1.45 16.38 ± 1.21
SAR at OEFS #2 19.86 ± 1.79 19.64 ± 1.47 20.57 ± 1.85 25.03 ± 2.15 18.39 ± 1.56 14.68 ± 1.23 22.3 ± 1.98 16.68 ± 1.41 12.65 ± 1.04
SAR at OEFS #3 21.51 ± 1.84 22.74 ± 2.01 23.46 ± 2.11 28.8 ± 2.39 13.53 ± 1.12 18.93 ± 1.61 27.6 ± 2.28 13.74 ± 1.14 20.3 ± 1.73
SAR at OEFS #4 27.68 ± 2.16 28.4 ± 2.17 25.42 ± 2.03 17.43 ± 1.29 13.18 ± 1.05 15.57 ± 1.15 19.32 ± 1.45 14.35 ± 1.02 13.82 ± 0.99

a, b = two same model Philips 3 T MRI systems, SD = standard deviation, TIRM = turbo inversion recovery magnitude, SPIR = spectral pre-saturation with inversion recovery, TSE = Turbo spin echo.