Fig. 4.
Alternating leaves of 5 mm in a 100 mm2 field detected by (a) the RL system and (b) the normalized profile along the centre of the image (line ‘L’ crossing the image a) in comparison with the profile obtained from the planned TPS test. Secondary axis from the plot is the difference between measured (film) and planned (TPS) profiles.