. 2020 Jun 11;14:67–73. doi: 10.1016/j.phro.2020.06.001

Table 2.

Form and parameters of dose rate regression functions.

Beam Frame Rate Gain Form Parameters
A B
6 MV
10 fps
18 dB Linear1 0.002 0.124
20 dB Linear 0.002 0.009
22 dB Linear 0.002 0.022
20 fps
20 dB * * *
22 dB * * *
6 MV FFF
10 fps
18 dB Exponential2 0.560 −2.634
20 dB Exponential 0.326 −1.210
22 dB Exponential 0.146 −0.163
20 fps
18 dB Linear 0.002 0.024
20 dB Linear 0.002 0.110
22 dB Exponential 0.879 −4.590
10 MV FFF
10 fps
18 dB Exponential 1.094 −5.904
20 dB Exponential 0.796 −4.024
22 dB Exponential 0.582 −2.728
20 fps
18 dB Linear 0.002 0.090
20 dB Linear 0.002 0.081
22 dB Linear 0.001 0.191

* Measurements could only be acquired at 600 MU/min. No regression function needed.

1

Linear regression functions were of the form: y=A×x+B.

2

Exponential regression functions were of the form: y=A×ln(x)+B.