Skip to main content
. 2020 Jun 11;14:67–73. doi: 10.1016/j.phro.2020.06.001

Table 3.

Root mean squared error of measurements acquired at non-reference conditions when standardized for gain and frame rate alone or with dose rate.

Standardized for gain and frame rate
Standardized for gain, frame rate, and dose rate
Beam Frame Rate Gain
Gain
18 dB 20 dB 22 dB 18 dB 20 dB 22 dB
6 MV
10 fps 22% 21% 24% 4% 2% 1%
20 fps * 0% 1% * 0% 1%
6 MV FFF
10 fps 29% 19% 22% 10% 8% 7%
20 fps 75% 68% 49% 3% 8% 10%
10 MV FFF
10 fps 91% 65% 43% 13% 17% 14%
20 fps 152% 140% 122% 5% 7% 11%

* Device was not sensitive enough to detect measurements at 18 dB and 20 fps.