Table 2.

Dose variation throughout the body during TSEBT in the standing and reclined positions.

TLD location Nr. of measurements Mean % of prescription dose
Standing Reclined P-value**
Top of head* 84 98 ± 21 80 ± 19 0.09
Forehead 84 101 ± 12 67 ± 13 ≤0.01
Behind right ear* 83 105 ± 16 68 ± 13 ≤0.01
Behind left ear* 81 106 ± 17 49 0.09
Sternal notch 87# 100 ± 10 58 ± 17 0.03
Right axilla+ 92# 69 ± 22 22 ± 31 ≤0.01
Left axilla+ 93# 69 ± 23 16 ± 24 ≤0.01
Mid chest* 93# 103 ± 11 80 ± 20 ≤0.01
Umbilicus 91# 108 ± 11 85 ± 22 ≤0.01
Right inguinal* 85 93 ± 18 78 ± 14 0.02
Left inguinal* 82 90 ± 21 77 ± 16 0.11
Right inner thigh* 84 92 ± 21 80 ± 25 0.3
Left inner thigh* 86# 88 ± 25 85 ± 29 0.8
Behind the right knee 84 103 ± 12 71 ± 7 ≤0.01
Behind the left knee 74 103 ± 13 83 ± 12 0.03
Upper back 93# 107 ± 9 85 ± 21 ≤0.01
Lower back* 91# 105 ± 7 92 ± 24 0.4
Perineum+ 74 20 ± 26 29 ± 19 0.2
Right sole+ 72 34 ± 20 71 ± 34 0.014
Left sole+ 62 31 ± 20 66 ± 38 0.1
Other regions 155
*

Sites affected by changing variables as demonstrated in Table 3.

**

P-value calculated with two sample T-test for 2 independent groups.

#

Sites may include doubled TLD measurements coming from the same patients

+

Sites require local boost radiations.

Number of total TLD measurements analyzed for each site in both positioning techniques.