. 2018 Dec 7;8:57–62. doi: 10.1016/j.phro.2018.10.003

Table 1.

Scanner models and acquisition parameters used in the 11 institutions participating in the Lungtech trial RTQA site visit.

Institute Scanner brand FOV Acquisition diameter (mm)# FOV Reconstruction diameter (mm) bin technique & kVp+ (V) Expos-ure (mA/s) Rotation speed (s) Slice thickness (mm) Scan length (mm) CTDI vol(mGy) Tube current (mA) Pitch *
A Toshiba Acquilion/LB 550 550 10, P 120 76 0.5 2 170 NA 250 0.12
170 NA 250 0.12
170 NA 250 0.075
B Philips Brillance big bore 600 480 10, A 120 600 0.5 2 162 NA 180 0.15
162 NA 132 0.11
162 NA 90 0.075
C Philips Gemini TF Big Bore 600 600 10, P 120 250 0.5 2 180 13.2 45 0.09
180 13.2 45 0.09
180 13.2 45 0.09
D Philips Brilliance big bore 600 568 10, A 120 801 0.44 3 240 NA 74 0.08
240 NA 74 0.041
240 NA 74 0.041
E Siemens Sensation open 500 500 8, P 120 400 0.5 1.5 79.5 NA 80 0.16
79.5 NA 80 0.1
120 NA 40 0.1
F Siemens sensation open 500 500 8, P 120 400 0.5 3 186 35.6 80 0.1
189 35.6 80 0.1
276 36.0 40 0.15
G Philips Brilliance big bore 600 568 10, A 120 801 0.44 3 171 47.4 74 0.08
171 47.4 132 0.041
171 47.4 90 0.041
H Philips Brilliance big bore 600 480 10, A 120 600 0.5 2 172 35.5 180 0.15
172 35.5 132 0.11
172 25.5 90 0.075
I Philips Brilliance big bore 600 500 10, P 120 350 0.5 2 226 20.7 105 0.15
226 20.7 53 0.11
226 20.7 77 0.08
J Toshiba Acquilion/LB 550 550 10, P 120 46 0.33 2 161 81.3 150 0.075
161 108.4 150 0.075
161 108.4 150 0.075
K Siemens Somatom Definition AS 500 500 10, P 120 80 0.5 3 176 NA 160 NA
176 NA 160 NA
176 NA 80 NA

*Where 3 values are given, different values are used for breathing periods of 3, 4 and 6 s, respectively. The pitch values that are larger than pmax are given in bold. NA: Not available. $Number of bins are given and type of binning: P = phase binning, A = amplitude binning, #FOV = field of view, +KVP = kilovolt peak