Mean and range deviations in volume (in % of the true volume) and amplitude (in cm) across institutions per sphere size and motion tested (minus signs mean an underestimation).

Motion Sphere radius (mm) Volume deviations (% of true volume)
Amplitude deviations (cm)
End of inspiration End of expiration Mid-ventilation Average CT
I: 15 mm/3 s 7.5 14.1% 2.2% 16.2% 43.2% −0.08
(−5.6, 40.8) (−8.4, 20.2) (3.3, 35.4) (22.1, 66.4) (−0.25, −0.01)
12.5 7.7% 1.1% 11.3% 41.4% −0.08
(−2.3, 33.9) (−1.6, 5.6) (3.5, 36.0) (26.3, 62.4) (−0.26, −0.01)
II: 15 mm/6 s 7.5 3.8% 1.0% 11.2% 12.4% −0.18
(−98.7, 39.7) (−1.5, 6.0) (−33.8, 29.9) (−33.8, 36.3) (−0.60, 0.01)
12.5 21.5% 2.1% 13.2% 31.8% −0.13
(−3.5, 68.5) (−1.5, 3.5) (0.1, 28.9) (13.7, 45.9) (−0.28, −0.01)
III: 25 mm/4 s 7.5 21.3% −1.0% 21.4% 22.8% −0.14
(−13.8, 64.8) (−12.9, 5.9) (−3.9, 51.1) (−2.1, 44.6) (−0.35, −0.02)