Skip to main content
. 2020 Nov 9;16:113–129. doi: 10.1016/j.phro.2020.09.011

Table 1.

Summary of empirically determined parameters for gamma analysis of in-vivo transit dosimetry results.

Normalization (Local/Global) Dose Difference Tolerance (%) Distance Tolerance (mm) Low Dose Threshold (%) Passing Tolerance Level (%)
Breast Local 7 6 20 90
Whole Brain RadioTherapy Local 7 3 20 90
Palliative treatments Local 7 5 20 93
H&N and Brain Global 3 3 20 95
Rectum Global 5 5 20 93
Other treatment sites with mask Global 5 3 20 95
Other treatment sites without mask (including lung, pelvis, abdomen,…) Global 5 5 20 95
Stereotactic 1 mm Local 10 1 20 95
Stereotactic 2 mm Local 10 2 20 95
Stereotactic 3 mm Local 10 3 20 95