Table 2.
Marker displacements between gold standard registrations and marker-based registrations.
Gold-std. registrations: Fully sampled MR Marker-based registrations: Breath-hold MR | Mean offset and standard deviation |
|||
---|---|---|---|---|
Phase Enc. (A/P) | Slice Sel. (S/I) | Readout (R/L) | ||
MR sequence (N = 16) |
T1w | 0.7 ± 0.3 | 0.1 ± 0.9 | 0.0 ± 0.3 |
T2*w | −0.4 ± 0.5 | 0.0 ± 0.6 | 0.1 ± 0.4 | |
T2w | 0.1 ± 0.3 | 1.0 ± 0.9 | 0.6 ± 0.5 | |
One way ANOVA result | p < 0.001 | p = 2.8 × 10−3 | p < 0.001 | |
Marker arrangement (N = 24) |
Coplanar | 0.3 ± 0.5 | 0.6 ± 0.9 | 0.3 ± 0.5 |
Non-coplanar | −0.1 ± 0.7 | 0.2 ± 0.9 | 0.1 ± 0.4 | |
2 sample t-test result | p = 0.04 | p = 0.13 | p = 0.07 | |
Test object orientation (N = 24) |
Horizontal | 0.1 ± 0.7 | 0.1 ± 0.6 | 0.0 ± 0.4 |
Vertical | 0.1 ± 0.5 | 0.7 ± 1.1 | 0.4 ± 0.4 | |
2 sample t-test result | p = 0.94 | p = 0.02 | p < 0.001 |
Results are listed with the p-value from the relevant statistical significance test performed. Here N is the number of measurements. Values shown in bold indicate results which reject the null hypothesis of a population mean equal to zero using a one-sample t-test (p < 0.05).