Skip to main content
. 2020 Oct 22;42(2):160–168. doi: 10.1097/MNM.0000000000001314

Table 4.

Summary of the overall heterogeneity index and the filtered heterogeneity index values for 81mKrypton and Technegas for all reconstructions

MEGP LEGP
Heterogeneity Index (±SEM) 81mKrypton Technegas P value 81mKrypton Technegas P value
ACSC
 Overall 1.03 ± 0.04 1.18 ± 0.07 0.006a 0.71 ± 0.03 1.24 ± 0.07 3·× 10−11a
 2–4 cm 0.38 ± 0.02 0.55 ± 0.03 3·× 10−7a 0.19 ± 0.01 0.57 ± 0.04 1·× 10−15a
AC
 Overall 0.98 ± 0.04 1.11 ± 0.06 0.009a 0.69 ± 0.03 1.17 ± 0.06 5·× 10−11a
 2–4 cm 0.34 ± 0.02 0.51 ± 0.03 2·× 10−7a 0.18 ± 0.01 0.53 ± 0.03 2·× 10−15a
NAC
 Overall 1.02 ± 0.04 1.08 ± 0.06 0.22 0.68 ± 0.02 1.12 ± 0.06 4·× 10−10a
 2–4 cm 0.37 ± 0.02 0.50 ± 0.03 1·× 10−5a 0.19 ± 0.01 0.50 ± 0.03 2·× 10−15a

AC, attenuation correction; ACSC, attenuation and scatter correction; LEGP, low energy general purpose; MEGP, medium energy general purpose; (N)AC, (non)attenuation corrected; SC, scatter corrected.

a

Indicate the significant differences between 81mKrypton and Technegas.