. 2021 Jan 15;100(2):e24064. doi: 10.1097/MD.0000000000024064

Table 2.

Characteristics of the high-intensity focused ultrasound ablation therapy.

Parameter HIFU alone (n = 30) CEUS+HIFU (n = 34) P
Sonication power (W) 399.4 ± 1.9 204.3 ± 37.6 <.001a
Treatment time (min) 181.3 ± 36.3 44.2 ± 15.4 <.001a
Sonication time (s) 2824.1 ± 419.8 358.4 ± 271.3 <.001a
Total energy used (kJ) 1127.8 ± 165.4 76.2 ± 58.8 <.001a
Energy efficiency factor (J/mm3) 10.9 ± 6.3 1.6 ± 0.3 <.001a
Non-perfused volume (cm3) 136.0 ± 73.8 161.6 ± 129.1 .872a
Fractional ablation (%) 70.0 ± 7.6 70.9 ± 13.5 .788a
Massive grayscale changes n, (%) 25/30 (83.3%) 34/34 (100%) .190b

Data are presented as the mean ± standard deviation or n/N (%). CEUS = contrast-enhanced ultrasonography, HIFU = high-intensity focused ultrasound, MGSC = massive grey scale change.