Table 2.

Surface dose (Ds) values (%) in the presence of air gaps under different thicknesses of gel bolus and water surface of radiation field analyzer-300, obtained from central axis depth dose curve of 6 MV clinical photon beam for different field sizes

Bolus thickness (cm) Field size (cm2) Air gap (cm) Reduction in “Ds”**

0.0 1.0 2.0 3.0
No bolus 5 × 5 48.7 9.9* - - - -
10 × 10 54.2 15.6* - - - -
15 × 15 59.9 21.6* - - - -
20 × 20 64.1 27.2* - - - -
25 × 25 67.7 32.4* - - - -
0.5 5 × 5 92.0 87.0 83.6 77.2 14.8
10 × 10 93.9 90.0 88.9 87.3 6.6
15 × 15 95.3 92.1 91.5 90.9 4.4
20 × 20 96.6 94.0 93.4 92.9 3.7
25 × 25 97.4 95.1 94.5 94.2 3.2
1.0 5 × 5 99.4 97.3 92.4 84.5 14.9
10 × 10 99.8 98.2 97.4 94.9 4.9
15 × 15 99.9 99.2 98.4 97.4 2.5
20 × 20 99.9 99.4 99.0 98.3 1.6
25 × 25 99.9 99.8 99.3 98.8 1.1
1.5 5 × 5 99.8 99.1 95.2 87.2 12.6
10 × 10 99.9 99.5 95.2 96.5 3.4
15 × 15 99.8 99.7 99.2 98.5 1.3
20 × 20 100.0 99.9 99.4 98.8 1.2
25 × 25 99.9 100.0 99.5 99.2 0.7

*Value after application of correction factor (Ci [L]),[5] **Value obtained by subtracting the Ds value without (0.0 cm) and with 3.0 cm air gap