Table 3.

Depth of dose maximum (dmax) values (cm) in the presence of air gaps between different thicknesses of gel bolus and water surface of radiation field analyzer- (RFA- 300), obtained from central axis depth dose curve of 6 MV clinical photon beam for different field sizes

Bolus thickness (cm) Field size (cm2) Air gap (cm) Shift of “Dmax”*

0 1.0 2.0 3.0
No bolus 5 × 5 1.4 - - - -
10 × 10 1.4 - - - -
15 × 15 1.4 - - - -
20 × 20 1.2 - - - -
25 × 25 1.2 - - - -
0.5 5 × 5 1.2 1.2 1.2 1.3 0.1
10 × 10 1.2 1.2 1.3 1.3 0.1
15 × 15 1.1 1.1 1.2 1.2 0.1
20 × 20 1.0 1.0 1.1 1.1 0.1
25 × 25 0.8 0.9 0.9 1.0 0.2
1.0 5 × 5 0.4 0.4 0.6 0.8 0.4
10 × 10 0.4 0.4 0.4 0.6 0.2
15 × 15 0.0 0.4 0.4 0.4 0.4
20 × 20 0.0 0.2 0.4 0.4 0.4
25 × 25 0.0 0.2 0.2 0.3 0.3
1.5 5 × 5 0.0 0.2 0.4 0.6 0.6
10 × 10 0.0 0.2 0.4 0.4 0.4
15 × 15 0.0 0.2 0.2 0.3 0.3
20 × 20 0.0 0.2 0.2 0.3 0.3
25 × 25 0.0 0.2 0.3 0.3 0.3

*Shift of dmax value obtained by subtracting the value of dmax without (0.0 cm) and with (3.0 cm) air gap