Skip to main content
. 2020 Oct 13;45(3):175–181. doi: 10.4103/jmp.JMP_53_20

Table 4.

Comparison of surface dose (Ds) values (%) of 6 MV photon beam for various field sizes cited in the literature with the present study

Study Field size (cm2)

5 × 5 8 × 8 10 × 10 10 × 20 15 × 15 20 × 20 25 × 25
Bilge et al.[16]a 10.0 * 15.0 * * 23.0 *
Akbas et al.[17]a 10.8 * 16.6 * * 28.1 *
Devic et al.[18]b 10.5 * 16.0 * 21.7 * *
Sigamani et al.[19]b 12.0 * 18.0 * 22.0 27.0 31.5
Butson et al.[20]a,b * 14.0 * 21.0 * * *
Ishmael Parsai et al.[21]c 10.5 * 16.0 * 21.5 * 31.5
Ishmael Parsai et al.[22]d 10.3 * 16.1 * 21.9 * 32.2
Apipunyasopon et al.[5]e 10.3 * 16.5 * 22.2 * 30.9
Present studyf 9.9 * 15.6 * 21.6 27.2 32.4

*Not quoted/measured. Measurements were done with following instruments/methods. aMarkus parallel plate chamber, bRadiochromic film, cExtrapolation chamber, dParallel plate chamber (readings applied with correction factor), eMonte Carlo simulation techniques, fCylindrical (CC13) chamber after applying the correction factor (“Ci [L]”)