Table 1.

Detailed Information of the Investigated Stents Including Stent Type, Material, Abbreviation (First Letters of Stent Name, Diameter/Length in mm), Number of Frequency Components Generated by the Stents and Relative Measurement Error of the Lumen Quantification (Calculated vs Nominal Lumen Diameter)

Stent Type Material Abbreviation Nr. Freq. Comp. SNR >5 Relative Measurement Error (%)
Biosensors, Biomatrix Neoflex 316L BiN 3/28 0 −3
Biosensors, BioFreedom 316L BiF 3.5/11 1 23
Boston Scientific, Taxus Liberté 316L TaL 4/38 0 −3
Boston Scientific, Taxus Liberté 316L TaL 5/32 0 −13
Boston Scientific, Express LD Vascular 316L ELV 7/57 0 −10
Boston Scientific, Express LD Vascular 316L ELV 10/37 0 −7
IDEV, Supera Nitinol Sup 4/40 0 −5
Gore, Tigris Nitinol Tig 5/40 0 −13
IDEV, Supera Nitinol Sup 5/60 0 −5
IDEV, Supera Nitinol Sup 6/40 1 0
Gore, Tigris Nitinol Tig 6/40 2 −9
Gore, Tigris Nitinol Tig 7/40 3 −10
Boston Scientific, Epic Nitinol Epi 7/99 0 −3
Gore, Tigris Nitinol Tig 8/40 2 −3
Boston Scientific, Promus Premier PtCr PrP 3/28 1 13
Boston Scientific, Promus Element Plus PtCr PEP 3/32 0 16
Boston Scientific, Synergy PtCr Syn 3/38 0 28
Boston Scientific, Promus Element Plus PtCr PEP 4/28 1 −3
Boston Scientific, Promus Premier PtCr PrP 4/28 0 −13
Boston Scientific, Rebel PtCr Reb 4/28 0 −3
Boston Scientific, Carotid Wallstent CoCr CaW 7/30 0 −4

Abbreviations: 316L, stainless steel; PtCr, platinum-chromium; CoCr, cobalt-chromium; SNR, signal-to-noise-ratio.