Skip to main content
. 2021 Jan 14;13:7–14. doi: 10.2147/OPTO.S280046

Table 3.

Results of the Regimen Test for CCP Used with GP Lenses

Lens Material Microorganism Initial CFU/Lens CFU/Lens and Soaking Solution After 6 Hours (n=12)
Silicone acrylate S. aureus 1.0 x 106 <1
P. aeruginosa 7.4 x 105 0
S. marcescens 6.3 x 105 <1
C. albicans 5.0 x 105 0
F. keratoplasticum 4.1 x 105 0
Fluorosilicone acrylate S. aureus 1.4 x 106 0
P. aeruginosa 8.4 x 105 0
S. marcescens 5.7 x 105 0
C. albicans 5.6 x 105 0
F. keratoplasticum 4.1 x 105 <1

Notes: Data reflect an average CFU from 4 lenses and 3 lots of solution for each lens type. Regimen test criteria is an average of <10 CFU per lens and soaking solution. For results of the soft (hydrophilic) and silicone hydrogel lenses tested, see Gabriel et al.7