Table 1.

Dimensions, gases, and operating parameters of the GC-IMS system

Parameters Value
Sample loop volume 250 μL
Sample loop gas flow 30 mLs/min
Sample gas Air (purified)
GC carrier gas flow 30 mLs/min
GC carrier gas Air (purified)
IMS drift gas flow 120 mLs /min
IMS drift gas Air (purified)
Dew point of drift, sample and GC carrier gas − 90 °C
IMS ionization source Tritium
IMS ionization source activity 130 MBq
IMS ionization source active diameter 10 mm
Field switching injection field 282.5 V/mm
Field switching blocking field − 0.25 V/mm and 0.45 V/mm
Drift length L 40.6 mm
Drift region diameter 8 mm
Drift region field 63.5 V/mm
Repetition rate 80 Hz
GC column length 1.07 m
Dew point of drift gas and sample gas − 90 °C
GC and sample loop temperature 40 °C
IMS temperature T 38 °C
Inner IMS pressure p 1001–1019.5 hPa