Effects of PIXI analysis on the variation between PBMC donors.

Basal OCR Basal GR
mean ± s.d. (pmol/min) CV (%) mean ± s.d. (mpH/min) CV (%)
Standard analysis 28.96 ± 5.11 17.7 6.38 ± 1.20 18.8
PIXI analysis 26.69 ± 3.74 14.0 5.94 ± 1.11 18.7
Effect of PIXI analysis − 2.27 ± − 1.37 − 3.7 − 0.44 ± − 0.09 − 0.1

PBMCs were isolated from sixteen healthy lean individuals (females, 18–25 years of age, BMI 18.5–25 kg/m2) and plated at 75–300 × 103 cells/well (n = 3–4) to generate a calibration curve that was used to transform total pixel intensity values into PIXI analyzed cell numbers. XF assay measurements were performed at 225 × 103 cells/well (n = 8) and basal OCR and basal GR levels were normalized against the number of plated cells (standard analysis) or PIXI analyzed cells (PIXI analysis). Here a summary of the changes in group mean, s.d. and CV for basal OCR and basal GR after standard analysis or PIXI analysis is represented.