Table 1.
Numerically simulated results of the transmit and local SAR efficiencies for different coils placed on a homogeneous human body phantom.
Simulation | Experiment | ||||
---|---|---|---|---|---|
ΔT | |||||
Coil | μT ⋅ W−1/2 | (μT ⋅ W−1/2 ⋅ kg1/2) | (W ⋅ kg−1) | (μT ⋅ W−1/2) | (K) |
LWA | 0.26 | 0.22 | 1.42 | 0.27 | 0.16 |
Dipole | 0.26 | 0.19 | 1.91 | 0.27 | 0.26 |
Stripline | 0.13 | 0.09 | 8.47 | – | – |
Loop | 0.07 | 0.05 | 5.25 | – | – |
Slot | 0.12 | 0.07 | 4.81 | – | – |
The field magnitude was taken at a depth of 7 cm.
Best values are highlighted in bold.