. 2021 Jan 19;12:455. doi: 10.1038/s41467-020-20708-w

Table 2.

Numerically simulated results of B1+ transmit and averaged 10 g-averaged SAR efficiencies for different coils in four-channel configurations placed around two different human body voxel models.

B1+ROI/Pacc B1+ROI/maxSAR10g maxSAR10g
Coil array Body model (μT ⋅ W−1/2) (μT ⋅ W−1/2 ⋅ kg1/2) (W/kg)
LWA Duke 0.091 0.193 0.22
Dipole Duke 0.085 0.142 0.37
Stripline Duke 0.068 0.075 0.81
Loop Duke 0.081 0.092 0.77
Slot Duke 0.054 0.049 1.1
LWA Gustav 0.105 0.214 0.24
Dipole Gustav 0.102 0.185 0.31

B1+ is averaged over the ROI, whereas the 10 g-averaged SAR is given in the maximum hotspot.