Table 2.

Sensitivity and Specificity of index relative to reference immunoasssays

Venous plasma samples
Sensitivity % (95% CI) Specificity % (95% CI)
Pos Neg
Manufacturer Positive Cut-off ≥ 1.0 Pos 8 0 44.4 (24.5, 66.2) 100 (97.9, 100)
Neg 10 177
Experimental Positive Cut-off ≥ 0.17 Pos 18 0 89.0 (67.2, 96.9) 100 (97.9, 100)
Neg 2 177

DBS samples compared to matched plasma samples tested on the Roche Elecsys ® Anti-SARS-CoV-2 immunassay.