SNM Settings at Initial (First) Programming Following IPG Implantation (Year 2017—European Expert Group)

Programming data 90 patients
Electrode configuration
Monopolar N (%) 11 (12)
Anode [+] adjacent to cathode [−] N (%) 46 (51)
Anode [+] most distant from cathode [−] N (%) 25 (28)
Longer (extended) cathode* N (%) 4 (4)
Other settings
Mean amplitude in V (± SD) 0.95 (± 0.5)
Mean pulse frequency/rate (Hz) 14.6
Mean pulse width (μs) 215
Cycling mode N (%) 6/90 (7)
*

A longer cathode can be obtained on selecting two adjacent electrodes as cathode [−].