Skip to main content
. 2021 Jan 6;21(2):330. doi: 10.3390/s21020330

Table 9.

AccuSet HIV-1 p24 Performance Panel (0800-0362).

Member DiaSorin LIAISON XL 1 Bio-Rad GenSys HIV1/2 PLUS O EIA 1 Avioq HIV-1 Microelisa 1 OraQuick ADVANCE 1 UCNP-LFA 2 Alere HIV Combo 3
0800-362-1 25.7 13.3 3.2 Pos 5.9 ++
0800-362-2 1.4 2.6 0.3 Neg 1.2 -
0800-362-3 0.2 0.8 0.2 Neg 0.8 -
0800-362-4 39.4 13.3 1.6 Neg 14.3 ++
0800-362-5 1.1 0.2 0.3 Neg 1.0 -
0800-362-6 0.3 0.8 0.3 Neg 0.4 -
0800-362-7 66.8 13.3 4.7 Pos 9.9 +++
0800-362-8 11.8 13.3 1.2 Neg 5.0 +
0800-362-9 0.2 0.2 0.3 Neg 0.5 -
0800-362-10 0.3 0.1 0.3 Neg 0.4 -
0800-362-11 5.0 5.3 0.3 Neg 0.6 -
0800-362-12 59.8 13.3 5.8 Pos 10.0 +++
0800-362-13 0.3 0.1 0.3 Neg 0.8 -

Numerical results are the S/Co ratios. 1 Results from the panel data sheet. S/Co ratios ≥ 1.0 are reactive. 2 For UCNP-LFA, S/Co ratios ≥ 1.0 are considered reactive. 3 Tests were run in our laboratory. Interpretation: +++ strong visible test line; ++ visible test line; + faint visible test line; +/- equivocal; - no visible test line.