Table 1.

Method validation parameters in melatonin-free matrix and real nail matrix. Calculated melatonin-free matrix parameters were linearity (in terms of weighting and range), LLOQ, LQC, MQC, HQC and LOD (in terms of accuracy and CV). Calculated real nail matrix parameters were within-run and total accuracy and CV; matrix effect (in terms of recovery and CV); the endogenous range of selected samples’ melatonin; and specificity of the method by comparing standard and samples. Units of range, level, LOD and concentration are fg/mg.

Melatonin-Free Matrix (CDER Regulations)
Linearity Weight. Range (fg/mg)
1/x 6.5–830
LLOQ LQC MQC HQC LOD
Level 6.5 13 0.104 830 3.5
Acc. (%) 113 115 104 103 -
CV (%) 8 6 4 4 -
Real nail matrix (Standard additions)
Within-run (n = 6) Total (n = 12) M.E. (n = 6) Endog Range
Acc. (%) 100 104 Rec. (%) 105 Conc. 30–110
CV (%) 10 13 CV (%) 7 (fg/mg)
Specificity
Standard Sample 1 Sample 2 Sample 3 Sample 4 Sample 5
Q/q ratio 0.22 0.23 0.21 0.24 0.23 0.22

Abbreviations: CDER: Center for Drug Evaluation and Research; Weight: weighting; LLOQ: lowest limit of quantification; LQC: lowest QC; MQC: medium QC; HQC: highest QC; LOD: limit of detection; Acc.: accuracy; CV: coefficient of variation; M.E.: matrix effect; Endog Range: endogenous range; Rec.: recovery; Conc: concentration; Q/q: ion ratio between quantitative transition (233 > 174) and qualitative transition (233 > 159).