. 2020 Dec 24;104:320–328. doi: 10.1016/j.ijid.2020.12.056

Table 2.

Healthcare workers’ assessment of respirators decontaminated in an oven. Each decontamination cycle was performed after the respirator had been used for a 12-h work shift in a ward dedicated to patients with coronavirus disease 2019 with 22 intensive care beds.

Assessment First decontamination cycle
Second decontamination cycle
Third decontamination cycle
Yes No n Yes No n Yes No n
Is the respirator dirty? 8 (24%) 25 (76%) 33 6 (33%) 12 (67%) 18 4 (44%) 5 (56%) 9
Is the elastic functional? 32 (91%) 3 (9%) 35 15 (83%) 3 (17%) 18 10 (91%) 1 (9%) 11
Is the metal nose clip functional? 21 (84%) 4 (16%) 25 9 (100%) 0 9 3 (75%) 1 (25%) 4
Functionality assessment Better Worse Same n Better Worse Same n Better Worse Same n
Does the respirator maintain a proper seal? 3 (9%) 7 (22%) 22 (69%) 32 1 (7%) 4 (26%) 10 (67%) 15 1 (10%) 4 (40%) 5 (50%) 10
How does breathing compare with previous use? 1 (3%) 7 (24%) 21 (73%) 29 0 2 (67%) 12 (86%) 14 0 3 (30%) 7 (70%) 10