Skip to main content
. 2020 Jun 17;129:104510. doi: 10.1016/j.jcv.2020.104510

Table 2.

Sensitivity and specificity of the PCR kits.

Company Target gene Sensitivity1 Specificity
Altona Diagnostics E 3.3 SARS-1
S 3.3 ND
Tibmolbiol E 25 SARS-1
RdRp 100 ND
ThermoFisher ORF1ab 33 ND
S 33 ND
N 3.3 ND
DAAN Gene ORF1ab 3.3 ND
N NA2, 3 NA2, 3
Kogene Biotech ORF1ab 100 ND
E 33 MERS, SARS-1
Liferiver ORF1ab 33 ND
N 3.3 SARS-1
Maccura Biotechnology ORF1ab 3.3 ND
E 3.3 SARS-1
N 3.3 ND
R-Biopharm E 50 ND
Sansure Biotech ORF1ab 3.3 ND
N 3.3 SARS-1
Sentinel Diagnostics B (RUO) E 33 SARS-1
RdRp 330 SARS-1
N 330 ND
Sentinel Diagnostics B (CEIVD) E 3.3 SARS-1
RdRp 330 SARS-1
N 330 SARS-1
Sentinel Diagnostics HK (RUO) ORF1ab 330 SARS-1
N 33 SARS-1
Sentinel Diagnostics HK (CE-IVD) ORF1ab 33 SARS-1
N 33 SARS-1
XABT ORF1ab 3.3 ND
E 3.3 ND
N 3.3 ND
Hecin Scientific RdRp 25 ND
N 25 ND
Reference assay E 3.3 SARS-1
RdRp 25 SARS-1

ND: Not detected; NA: not available.

1

Expressed as lowest amount of RNA copies detectable in 3/3 replicates.

2

For N-gene assay only background signal was observed.

3