. 2020 Dec 17;289:114042. doi: 10.1016/j.jviromet.2020.114042

Fig. 1.

Fig. 1

below shows box plots comparing the cycle threshold, Ct using the LDT Light Cycler 480 [n = 9] when specimens were extracted using MagNAPure 96 extraction versus when specimens were extracted using EpMotion 5075/Geneaid Viral RNA extraction kit. Nine specimens extracted on MP 96 extraction with a median Ct of 26.6 (interquartile range, IQR 23.3-28) had a higher median Ct of 31.6 (IQR 29.1–33.8) when extracted on the EpMotion 5075/Geneaid Viral RNA extraction kit and amplified using the LDT on the LC 480.