Table 1.

Scan parameters of 2D CSE-MRI and 3D CSE-MRI in the phantom and animal studies at 3.0T and the preliminary human studies at 1.5T.

# TE TEinit/ΔTE (ms) TR (ms) FA (°) FOV (mm3) ACQ Resolution (mm3) NOA ACQ BW (kHz) PI Acceleration Phase/Slice Scan Time (minutes) Temporal Footprint per Slice
Phantom Study (3.0T)
 2D CSE-MRI 16 1.8/1.0 18.9 9 220×220×69+ 2.3×2.3×2.3+ 1 ±100 1.5 / 1.0 1:34 3.1 seconds
 2D CSE-MRI high SNR 16 1.8/1.0 18.9 9 220×220×72+ 2.3×2.3×3.0+ 4 ±62.5 1.5 / 1.0 6:13 15.5 seconds
 3D CSE-MRI× 8 1.6/1.3 12.8 9 220×220×68 1.5×1.5×1.2 1* ±62.5 1.5 / 1.0 1:40 1:40 minutes
 3D CSE-MRI high SNR× 8 1.6/1.3 12.8 9 220×220×72 1.5×1.7×8.0 1* ±143 1.5 / 1.5 0:38 0:38 minutes
Animal Study (3.0T)
 2D CSE-MRI 16 1.8/1.0 18.7 9 180×160×140+ 1.9×1.8×3.0+ 1 ±62.5 1.5 / 1.0 2:11 2.8 seconds
 3D CSE-MRI 8 1.6/1.3 12.9 9 160×140×140 1.1×1.0×2 1* ±62.5 1.5 / 1.0 4:54 4:54 minutes
Human Study (1.5T)
 2D CSE-MRI 12 1.5/0.8 12.7 10 420×378×224+ 2.9×2.9×4.0 1 ±62.5 1.5 / 1.0 2:14 2.4 seconds
 3D CSE-MRI 6 1.2/2.0 13.9 9 420×378×228 2.2×2.2×3 1* ±83.3 2.0 / 1.5 1:48 1:48 minutes
 T2 SSFSE 1 100/0 4000 155 380×380×210 1.5×1.5×5 1 ±83.3 2.0 / 1.0 2:40 2.8 seconds
+

Contiguous slices with slice gap=0 mm;

*

3D corner cutting used (20% k-space reduction).

×

No respiratory gating used in phantom study.

Scan time with respiratory gating. CSE: chemical shift-encoded; FA: flip angle; FOV: field of view; ACQ: acquisition; NOA: number of averages; BW: sampling bandwidth; PI: parallel imaging; SSFSE: single-shot fast spin-echo