Table 4.

Repeatability analysis for placental R2* measured in 2D and 3D CSE-MRI acquisitions in human subjects at 1.5T.

MDwithin (s−1) SDwithin (s−1) CR (s−1)
2D CSE-MRI All (N=21) 0.7 3.4 9.5
Superior (N=7) 1.0 2.7 7.6
Middle (N=7) 1.6 3.4 9.4
Inferior (N=7) −0.7 4.1 11.3
3D CSE-MRI All (N=21) −0.2 3.8 10.6
Superior (N=7) 0.5 5.1 14.1
Middle (N=7) −0.4 2.4 6.6
Inferior (N=7) −0.6 4.1 11.3

Note – CSE: chemical shift-encoded. MDwithin=within-technique mean difference. SDwithin=within-technique standard deviation. CR=coefficient of repeatability