Table 1.
ANTERION | CASIA2 | IOLMaster 700 | ANTERION vs IOLMaster 700 | CASIA2 vs IOLMaster 700 | ANTERION vs CASIA2 | ||||
---|---|---|---|---|---|---|---|---|---|
ICC | Difference | ICC | Difference | ICC | Difference | ||||
Anterior average K, D | 43.92 ± 1.62 | 44.02 ± 1.57 | 43.98 ± 1.63 | 0.991 | − 0.059 ± 0.213 | 0.989 | 0.038 ± 0.233 | 0.990 | − 0.097 ± 0.201 |
Anterior steep K, D | 44.42 ± 1.62 | 44.51 ± 1.55 | 44.38 ± 1.60 | 0.974 | 0.034 ± 0.367 | 0.959 | 0.13 ± 0.439 | 0.962 | −0.096 ± 0.433 |
Anterior flat K, D | 43.42 ± 1.70 | 43.53 ± 1.66 | 43.58 ± 1.68 | 0.976 | −0.166 ± 0.334 | 0.978 | −0.059 ± 0.348 | 0.970 | −0.107 ± 0.403 |
Anterior cylinder K, D | 1.01 ± 0.71 | 0.99 ± 0.76 | 0.80 ± 0.46 | 0.598 | 0.21 ± 0.516 | 0.443 | 0.189 ± 0.653 | 0.526 | 0.021 ± 0.719 |
Total average K, D | 43.33 ± 1.63 | 42.89 ± 1.52 | 43.94 ± 1.61 | 0.922 | −0.61 ± 0.255 | 0.808 | − 1.049 ± 0.237 | 0.952 | 0.439 ± 0.232 |
Total steep K, D | 43.85 ± 1.64 | 43.36 ± 1.57 | 44.37 ± 1.64 | 0.920 | −0.525 ± 0.425 | 0.801 | − 1.016 ± 0.443 | 0.918 | 0.491 ± 0.453 |
Total flat K, D | 42.81 ± 1.69 | 42.42 ± 1.56 | 43.51 ± 1.62 | 0.895 | −0.708 ± 0.359 | 0.790 | − 1.094 ± 0.353 | 0.938 | 0.386 ± 0.438 |
Total cylinder K, D | 1.04 ± 0.74 | 0.94 ± 0.73 | 0.86 ± 0.47 | 0.505 | 0.183 ± 0.602 | 0.441 | 0.078 ± 0.649 | 0.461 | 0.105 ± 0.761 |
ALa, mm | 24.00 ± 1.60 | NA | 23.99 ± 1.54 | 0.999 | −0.005 ± 0.053 | NA | NA | NA | NA |
CCT, μm | 535.89 ± 34.29 | 533.60 ± 34.26 | 535.19 ± 35.04 | 0.994 | 0.702 ± 3.735 | 0.986 | −1.596 ± 5.57 | 0.981 | 2.298 ± 6.304 |
ACD, mm | 2.62 ± 0.39 | 2.60 ± 0.48 | 2.53 ± 0.47 | 0.990 | 0.058 ± 0.033 | 0.982 | 0.075 ± 0.052 | 0.994 | − 0.015 ± 0.049 |
LTb, mm | 4.63 ± 0.43 | 4.60 ± 0.55 | 4.54 ± 0.52 | 0.915 | 0.154 ± 0.159 | 0.963 | 0.088 ± 0.123 | 0.961 | 0.064 ± 0.141 |
Continuous data are presented as mean ± standard deviation
Differences between two devices were defined by subtracting the latter value from the former value
ACD anterior chamber depth, AL axial length, CCT central corneal thickness, D Diopter, ICC intraclass correlation coefficient, K keratometry, LT lens thickness, NA not available, vs versus
aAL was analyzed with 41 eyes that could be measured by both ANTERION and IOLMaster 700
bLT was analyzed with 44 eyes that could be measured by all three devices