Table 2.
Performance of compact methanol detectors
| Type | Reference | LOQa (vol%) | Analysis time (s) | Methanol selectivityb |
Reusablec | Stabilityd (days) | Validated with hand sanitizers | Price (USD) | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Ethanol | 1-propanol | 2-propanol | Butanone | Glycerol | ||||||||
| Chemoresistive | Guo et al., (2011) | 0.02 (g) | 0.5 | ✓ | ||||||||
| Zhao et al., (2012) | 8∙10−7 (g) | <2 | ✓ | |||||||||
| This work | 0.01 (L) 10−4 (g)g |
≤90 | ∞ | ∞ | ∞ | ∞ | ∞ | ✓ | 107i | ✓ | ||
| ECe | Ou et al., (2019) | 0.15 (L) | 260 | ✓ | ||||||||
| Meng et al., (2020) | 2∙10−4 (g) | 60 | ~1 | ✓ | 15 | |||||||
| Optical | DX4015 (Gasmet Technol.) | 3∙10−4 (g) | <120 | ✓ | >10′000 | |||||||
| Spectrum Two FT-IR Spectrometer (PerkinElmer) | 0.03 (L) | 30 | ✓ | ✓ | ||||||||
| Huang et al., (2018) | 4 (L) | <2 | 0.7h | ✓ | ||||||||
| CMf | Alert for Methanol (Neogen) | 0.35 (L) | 600 | single use | 20 (per analysis) | |||||||
lowest gas- (g) or liquid- (L) phase concentration measured.
highest ratio of response methanol vs. response confounder.
repeated use of same detector/reagent.
stability during repeated measurements without significant performance loss.
electrochemical.
colorimetric.
data from van den Broek et al. (2019).
authors suggest ethanol vs. methanol discrimination through different sensor recovery times.
data from Abegg et al. (2020).