Skip to main content
. 2021 Jan 12;24(2):102050. doi: 10.1016/j.isci.2021.102050

Table 2.

Performance of compact methanol detectors

Type Reference LOQa (vol%) Analysis time (s) Methanol selectivityb
Reusablec Stabilityd (days) Validated with hand sanitizers Price (USD)
Ethanol 1-propanol 2-propanol Butanone Glycerol
Chemoresistive Guo et al., (2011) 0.02 (g) 0.5
Zhao et al., (2012) 8∙10−7 (g) <2
This work 0.01 (L)
10−4 (g)g
≤90 107i
ECe Ou et al., (2019) 0.15 (L) 260
Meng et al., (2020) 2∙10−4 (g) 60 ~1 15
Optical DX4015 (Gasmet Technol.) 3∙10−4 (g) <120 >10′000
Spectrum Two FT-IR Spectrometer (PerkinElmer) 0.03 (L) 30
Huang et al., (2018) 4 (L) <2 0.7h
CMf Alert for Methanol (Neogen) 0.35 (L) 600 single use 20 (per analysis)
a

lowest gas- (g) or liquid- (L) phase concentration measured.

b

highest ratio of response methanol vs. response confounder.

c

repeated use of same detector/reagent.

d

stability during repeated measurements without significant performance loss.

e

electrochemical.

f

colorimetric.

h

authors suggest ethanol vs. methanol discrimination through different sensor recovery times.

i

data from Abegg et al. (2020).