Skip to main content
. 2020 Sep 6;35(1):e23563. doi: 10.1002/jcla.23563

TABLE 2.

The results of the LAMP assay compared to those of the RUT and PCR and sequencing analysis for the diagnosis of H pylori and the detection of the A2143G and T2182C point mutations in 23S rRNA

Development cohort RUT PCR
LAMP for H pylori a Negative Positive Statistics b Negative Positive Statistics
Negative 343 5 P = .017 348 0 P = 1.0
Positive 17 229 κ = 0.923 0 246 κ = 1.0
LAMP for 2143G or 2182C Sequencing A2143G Sequencing T2182C
2143A 2143G Statistics 2182T 2182C Statistics
2143G Negative 182 0 P = 1.0
Positive 0 64 κ = 1.0
2182C Negative 30 0 P = 1.0
Positive 0 216 κ = 1.0
a

DNA used in LAMP analysis was extracted in a laboratory setting with a QIAamp® DNA Mini Kit (QIAGEN, #51306). Twenty‐eight patients with an extracted DNA concentration of <50 ng/µL were excluded from the PCR and LAMP analyses.

b

P from McNemar test and κ from unweighted pairwise Cohen's Kappa.