Table 2.
FRETS‐VWF73 assay | HemosIL AcuStar ADAMTS13 Activity assay | |
---|---|---|
Level of automation | Highly manual | Fully automated |
Substrate | FRETS‐VWF73 | GST‐VWF73 |
Detection principle | Fluorescence | Chemiluminescence |
Limit of detection | 10 IU/dL | 0.2 IU/dL |
Contamination | Not applicable | None |
Intra‐assay precision (CV) | 6.9% | 5.6% |
Interassay precision (CV) | 10.2% | 9.7% |
Assay length a | 1 hour | 33 minutes b |
Abbreviations: CV, coefficient of variation.
This duration does not include sample preparation, calibration and data analysis.
This is the time that the HemosIL AcuStar ADAMTS13 activity assay needs to assess the first sample; each subsequent sample will take an extra minute.