Skip to main content
. 2020 Dec 15;5(1):81–93. doi: 10.1002/rth2.12461

Table 5.

Diagnostic performance for acute phase of TTP of the HemosIL AcuStar ADAMTS13 Activity assay in 539 plasma samples

FRETS‐VWF73
<10 IU/dL (n = 142) ≥ 10 IU/dL (n = 397)
HemosIL AcuStar ADAMTS13 Activity assay <10 IU/dL (n = 129) 128 1
≥ 10 IU/dL (n = 410) 14 396
Sensitivity (%) 90.1
Specificity (%) 99.7
Positive predictive value (%) 99.2
Negative predictive value (%) 96.6
Cohen’s kappa coefficient 0.93