Skip to main content
. 2021 Jan 29;16(1):e0245970. doi: 10.1371/journal.pone.0245970

Fig 3.

Fig 3

Linear regression (a) and Bland-Altman (b) plots comparing in vivo MRF T1 values obtained from all tissues combined at 1.5T and 3T systems. On Fig (b), dotted lines represent upper and lower 95% limits of agreement and bold lines represent the mean biases with appropriate captions included. MRF = magnetic resonance fingerprinting, SD = standard deviation.