Table 1.
Automated Sensor Feature | Condition (Based on Programmed Thresholds) | Required for Therapy Delivery | Used for Stimulation Amplitude Modulation |
---|---|---|---|
Pitch | Recumbent | X | X |
Activity | Inactive | X | X |
Time of day | Time within typical sleep window | X | X |
Lead impedance | Impedance within an acceptable range | X | |
Respiration | Transthoracic impedance measurements within a preset range | X | |
Body position | Position tracked within quadrant angles (supine, left, right, prone) | X | |
Week number | Stimulation amplitude increased over several weeks allows for acclimatization | X |
Once required conditions are met, the device automatically increases or decreases stimulation current levels based on information about respiration, body pitch/position, activity, and/or time on therapy.