Table 1.

Technical requirements, performance, and costs/availability for motion correction methods.

METHOD System Requirements Motion tracking
(temporal and spatial resolution)
B0 field tracking
(temporal and spatial resolution)
Performance, costs and availability
Retrospective: Post-processing Custom post-processing software No localization update Temporal resolution is determined by the TR of MRS sequence Lower cost of software solution
Prospective: Navigators Custom MR sequence, on-line reconstruction and real-time feedback 0.05–0.1 mm & 0.05–0.1 deg tracking accuracy, 300–1000 ms update rate 0.05–0.1 mm & 0.05–0.1 deg tracking accuracy, 300–1000 ms update rate Lower cost of software solution
Prospective: External tracking (optical and NMR probes Additional hardware and software, integration with MR scanner 0.005–0.1 mm & 0.003–0.1 deg tracking accuracy, 12–30 ms update rate No field update is possible without additional frequency navigators Higher costs for commercial and custom-made systems
Prospective: NMR probes Additional hardware and software, integration with MR scanner 0.01–0.1 mm & 0.01–0.1 deg tracking accuracy, 300–1000 ms update rate Field is measured outside the head, and computational EM modelling can approximate the internal field with limited accuracy. Higher costs for commercial and custom-made systems