Table 1.
METHOD | System Requirements | Motion tracking (temporal and spatial resolution) |
B0 field tracking (temporal and spatial resolution) |
Performance, costs and availability |
---|---|---|---|---|
Retrospective: Post-processing | Custom post-processing software | No localization update | Temporal resolution is determined by the TR of MRS sequence | Lower cost of software solution |
Prospective: Navigators | Custom MR sequence, on-line reconstruction and real-time feedback | 0.05–0.1 mm & 0.05–0.1 deg tracking accuracy, 300–1000 ms update rate | 0.05–0.1 mm & 0.05–0.1 deg tracking accuracy, 300–1000 ms update rate | Lower cost of software solution |
Prospective: External tracking (optical and NMR probes | Additional hardware and software, integration with MR scanner | 0.005–0.1 mm & 0.003–0.1 deg tracking accuracy, 12–30 ms update rate | No field update is possible without additional frequency navigators | Higher costs for commercial and custom-made systems |
Prospective: NMR probes | Additional hardware and software, integration with MR scanner | 0.01–0.1 mm & 0.01–0.1 deg tracking accuracy, 300–1000 ms update rate | Field is measured outside the head, and computational EM modelling can approximate the internal field with limited accuracy. | Higher costs for commercial and custom-made systems |