Table 2.

Recommendations for the technical performance of prospective real-time motion and shim correction for various required levels of precision in MRS. Note: 1) precision has been derived for either pure translation or rotation considering the case of an isotropic SVS voxel; 2) for MRSI, similar precision thresholds are hard to derive theoretically and the acceptable performance level for motion correction should be empirically evaluated by test-retest. An SVS voxel of 10 mm × 10 mm × 10 mm and anterior-posterior brain diameter of 220 mm were considered. Translation for any voxel and rotation angle for a frontal pole voxel, which is displaced the most by rotation around an axis through the center of the atlanto-axial joint are calculated according to the formulas given above. All reported real-time methods have higher translation and rotation precisions than the values listed in the table. The guidelines for homogeneity of the B0 field are the same as recommended in the B0 shimming consensus paper.3

Basic Moderate Advanced
20% error in estimation of metabolite concentration
  • 0.7 mm translation

  • 1.0° rotation

  • frequency update

10% error in estimation of metabolite concentrations
  • 0.4 mm translation

  • 0.5° rotation

  • frequency update

  • shim update

5% error in estimation of metabolite concentrations
  • 0.2 mm translation

  • 0.3° rotation

  • frequency update

  • shim update

  • reacquisition