Table 1.

The accuracy and sensitivity of a bioluminescent signal-based digital LAMP assay.

Target RNA Concentrations
(Copies/mL)
Target RNA Concentrations in Reaction Mix (Copies/μL) (10 μL Reaction Mixtures) Theoretical Calculation
(N = 20,000)
Actual Results
λ Np Nn/N Np N λ C (Copies/μL)
104 1 0.0005 10 0.9995 9 18,554 0.00049 0.98
105 10 0.0050 99 0.9950 77 17,386 0.00443 8.89
106 100 0.0500 969 0.9516 669 19,097 0.03503 70.06

N: the total number of droplets. Np: number of positive droplets. Nn: number of negative droplets. λ: copies per droplet = −ln(Nn/N) = Copies in the reaction/Total droplets = Concentration in reaction (Copies/μL) × V (volume of each droplet). V = Reaction volume/N = 0.0005 μL. P(0) = Nn/N = e−λ, λ = −ln(Nn/N) = Concentration in reaction × V. Concentration in reaction = [−ln(N/N)]/V. Target RNA concentration = 10 × CCopies/μL = 1043 × CCopies/mL. P(0) = Nn/N = e−λ, λ = −ln(Nn/N) = CCopies/mL × V.