Table 2.

The accuracy of the bioluminescent signal-based simulated digital LAMP assay.

Target RNA Concentrations
(Copies/mL)
Target RNA Concentrations in Reaction (Copies/μL)
(1000 μL Reaction Mixtures)
Theoretical Calculation
(N = 96)
Actual Results
λ Np Nn/N Np N λ C (Copies/μL)
10 0.001 0.0104 1 0.9896 2 96 0.0208 0.0019
102 0.01 0.1040 10 0.9019 8 96 0.0870 0.0084
103 0.1 1.0416 62 0.3535 58 96 0.9268 0.0885
104 1 10.4167 96 0 96 96 ≥9.6 ≥0.96

N: the total number of wells. Np: number of positive wells. Nn: number of negative wells. λ: copies per well = −ln(Nn/N) = Copies in the reaction/Total wells = Concentration in reaction (Copies/μL) × V (volume of each well). V = Reaction volume/N = 10.4 μL. P(0) = Nn/N = e−λ, λ = −ln(Nn/N) = Concentration in reaction × V. Concentration in reaction = [−ln(N/N)]/V. Target RNA concentration = 10 × CCopies/μL = 104 × CCopies/mL.