Results of accuracy evaluation for normal-phase HPTLC technique (mean ± SD; n = 6).

Conc. (ng Band−1) Conc. Found (ng Band−1) ± SD Recovery (%) RSD (%)
10 10.14 ± 0.10 101.40 0.98
500 496.54 ± 2.67 99.30 0.53
1000 989.78 ± 3.85 98.97 0.38