Comparison of different amperometric glucose biosensors.

Glucose Biosensor Linear Range (mM) LOD (mM) KMapp. (mM) Reference
GR/PANI:rGO/GOx 0.5–50 0.089 74.4 Current work
SPCE/MWCNTs:AuNP:rGO:/GOx 1–10 0.064 0.74 [24]
GR/rGO/PANI/GOx 0.1–8.5 0.0007 1.52 [25]
CC/PANI/GOx 1–8 0.050 - [46]
Au/PANI:PAA/GOx 0.1–1.3 0.024 - [47]
GCE/G:GOx 0.1–10 0.010 - [48]
GCE/rGO:C60/GOx 0.1–12.5 0.035 4.44 [49]

Note: SPCE, screen-printed carbon electrode; MWCNTs multiwall carbon nanotubes; AuNP, gold nanoparticle; CC, carbon cloth; PAA, poly(acrylic acid); GCE, glassy carbon electrode; G, graphene; C60, fullerene-C60.