HPLC instrumentation and parameters.

Analyte(s) Vicine, convicine, divicine, and isouramil in FBPI Vicine in NP Levodopa Tyramine
LC system Agilent Model 1260 (Agilent Technologies, Wilmington, DE, USA)
Detector Agilent G4212B diode array Agilent G1321A fluorescence
Detection wavelengths 276 nm, 214 nm, 273 nm, 278 nm, 283 nm 276 nm, 214 nm Ex = 280 nm
Em = 310 nm
Ex = 275 nm
Em = 303 nm
LC column YMC-Pack ODS-AQa Zorbax Eclipse Plus C18b Zorbax Eclipse Plus C18b YMC-Pack ODS-AQa
Temperature 25°C 15°C 20°C 15°C
Mobile phase A 0.05 M KH2PO4, pH 2.9 1000/25 (v/v) 0.05 M KH2PO4, pH 2.9/ACN 0.02 M KH2PO4, pH 2.9
Mobile phase B 200/800 (v/v) H2O/ACN
Flow rate 0.5 mL/min 0.4 mL/min 0.4 mL/min 0.5 mL/min
Injection 10 μL 10 μL 5 μL 10 μL
Elution program 0% B 0-20 min, 100% B 20-25 min, 0% B 25-45 min (end) 0% B 0-25 min, 100% B 25-30 min, 0% B 30-45 min (end) 0% B 0-25 min, 100% B 25-30 min, 0% B 30-50 min (end) 0% B 0-5 min, 0-10% B 5-25 min (linear), 100% B 25-30 min, 0% B 30-45 min (end)