. 2021 Feb 8;197:113956. doi: 10.1016/j.jpba.2021.113956

Table 2.

The effect of the PL removal step. Calibrants were prepared in human plasma, porcine plasma and water at the concentrations of 0.1, 2, 50, 100, 150, 200, 250 and 300 ng/mL. Linear calibration curves with concentration as an independent variable (y = ax + b) were created by weighted (1/x) regression analysis of the IS normalised peak areas. The average of two experiments is presented.

Calibrant matrix type PL removal included IS a (± SD) b (± SD) R2
Human plasma Yes THC-OH D3 0.0321 ± 0.0006 0.070 ± 0.096 0.998
Porcine plasma Yes THC-OH D3 0.0321 ± 0.0012 0.152 ± 0.172 0.998
Water Yes THC-OH D3 0.0324 ± 0.0005 0.111 ± 0.086 0.999
Human plasma No THC-OH D3 0.0251 ± 0.0003 0.040 ± 0.048 0.999
Porcine plasma No THC-OH D3 0.0321 ± 0.0006 0.132 ± 0.103 0.998
Water No THC-OH D3 0.0364 ± 0.0005 0.073 ± 0.075 0.999
Human plasma Yes Diazepam D5 0.0583 ± 0.0006 0.131 ± 0.099 0.999
Porcine plasma Yes Diazepam D5 0.0584 ± 0.0015 0.153 ± 0.254 0.996
Water Yes Diazepam D5 0.0588 ± 0.0008 0.091 ± 0.088 0.999
Human plasma No Diazepam D5 0.0355 ± 0.0005 0.063 ± 0.078 0.999
Porcine plasma No Diazepam D5 0.0464 ± 0.0005 0.097 ± 0.084 0.999
Water No Diazepam D5 0.0538 ± 0.0007 0.061 ± 0.121 0.999