Table 2.
The effect of the PL removal step. Calibrants were prepared in human plasma, porcine plasma and water at the concentrations of 0.1, 2, 50, 100, 150, 200, 250 and 300 ng/mL. Linear calibration curves with concentration as an independent variable (y = ax + b) were created by weighted (1/x) regression analysis of the IS normalised peak areas. The average of two experiments is presented.
| Calibrant matrix type | PL removal included | IS | a (± SD) | b (± SD) | R2 |
|---|---|---|---|---|---|
| Human plasma | Yes | THC-OH D3 | 0.0321 ± 0.0006 | 0.070 ± 0.096 | 0.998 |
| Porcine plasma | Yes | THC-OH D3 | 0.0321 ± 0.0012 | 0.152 ± 0.172 | 0.998 |
| Water | Yes | THC-OH D3 | 0.0324 ± 0.0005 | 0.111 ± 0.086 | 0.999 |
| Human plasma | No | THC-OH D3 | 0.0251 ± 0.0003 | 0.040 ± 0.048 | 0.999 |
| Porcine plasma | No | THC-OH D3 | 0.0321 ± 0.0006 | 0.132 ± 0.103 | 0.998 |
| Water | No | THC-OH D3 | 0.0364 ± 0.0005 | 0.073 ± 0.075 | 0.999 |
| Human plasma | Yes | Diazepam D5 | 0.0583 ± 0.0006 | 0.131 ± 0.099 | 0.999 |
| Porcine plasma | Yes | Diazepam D5 | 0.0584 ± 0.0015 | 0.153 ± 0.254 | 0.996 |
| Water | Yes | Diazepam D5 | 0.0588 ± 0.0008 | 0.091 ± 0.088 | 0.999 |
| Human plasma | No | Diazepam D5 | 0.0355 ± 0.0005 | 0.063 ± 0.078 | 0.999 |
| Porcine plasma | No | Diazepam D5 | 0.0464 ± 0.0005 | 0.097 ± 0.084 | 0.999 |
| Water | No | Diazepam D5 | 0.0538 ± 0.0007 | 0.061 ± 0.121 | 0.999 |