Table 3.

Comparison of tissue conductivities measured using a four-probe method by an impedance analyzer and average conductivity measured in images over phantom regions of interest including chicken, potato or agar materials (S/m) using the proposed method and horizontal (‘H’) or vertical (‘V’) current injection or the non-iterative harmonic Bz method and full or half NEX data.

Material Conductivity (S/m)
Method Type Chicken Potato Agar
Impedance Analyzer - 0.70 ± 0.0169 0.18 ± 0.006 0.75 ± 0.018
Proposed H 0.69 ± 0.0189 0.59 ± 0.0133 0.76 ± 0.0499
V 0.64 ± 0.0174 0.54 ± 0.0141 0.72 ± 0.0313
Non-iterative Harmonic Bz Full NEX 0.64 ± 0.0156 0.54 ± 0.0083 0.74 ± 0.0181
Half NEX 0.63 ± 0.0391 0.50 ± 0.0610 0.77 ± 0.1099