Experimental setup and
confirmation of the molecular identity of
the target compound. Three sections of the study tissue were placed
on a slide. One served as control, one was sprayed with three differently
labeled versions (L1, L2, and L3) of the target compound, each at
a different concentration (C1–C3), and a third tissue section
was spotted with a seven-level (C1′–C7′) dilution
series of L4. Finally, a fourth label (L4) was applied at the same
concentration as L2 (C2) on both adjacent tissue sections for normalization
purposes. The tissue studied is a pig colon tissue of which a hematoxylin
and eosin (H&E) staining is shown in (b, left) and the MALDI-MSI
image of the target histone H4 peptide is shown in (b, right). The
identification of the target peptide was performed on tissue using
high-mass accuracy delivered by a Q-exactive Orbitrap mass spectrometer
with a mass error of 0.75 ppm (c), by a comparison of the observed
and the expected isotopic patterns (d), and MS/MS experiments (Supporting Information Figure 1).