Skip to main content
. 2021 Feb 8;10(2):5. doi: 10.1167/tvst.10.2.5

Table 2.

Summary of Commercially Available Two-Wavelength Retinal Oximetry Systems118

Name (Manufacturer) Components Fundus Camera Integrated Cameras Light Wavelengths Mean Retino-Vascular OS (SD) Minimum Vessel Diameter Additional Functions
Oxymap T1 (Oxymap ehf, Reykjavik, Iceland) Fundus camera, two digital cameras, Oxymap Analyzer software TRC-50DX (Topcon Corporation, Tokyo, Japan) Insight IN1800, 1600 × 1200 square pixels (Diagnostic Instruments Inc., Sterling Heights, MI) 600 nm and 570 nm Arterioles: 98% (3.9%) Venules: 65% (4.1%) 74 µm
Dynamic Vessel Analyzer (Imedos GmbH, Jena, Germany) Fundus camera, digital camera, RVA software FF450 (Carl Zeiss Meditec, Jena, Germany) HV-C20A (Hitachi, Ltd., Tokyo, Japan) 610 nm and 548 nm Arterioles: 92% (2.7%) Venules: 55% (5.6%) 90 µm Static vessel analyzer, vessel dilatation with flicker